Class I Test Flowchart
Start
#
CAUTION
Short-circuit between L/N and PE is doubt when "no" and "OFF"
appears on the LCD in turns at the insulation resistance measurement
of the Extension leads test for Class I & II while DUT is on. Testing
should be immediately suspended.
Follow the procedure described in 5.2-3 and do NULL setting before a
measurement.
Alligator clip must make good contact w
ith the enclosure of the DUT.
When the terminal is open or the resistance value exceeds measuring
range, OL mark (over range display) appears on the LCD.
Do not touch the device under test while testing is in progress. Since
a high voltage of 500V, user may get electrical shock.
. Protective conductor resistance.
RPE
<
=
1Ω?
.Appliance switch test
(3). Insulation resistance between L/N and PE.
LnE
>
=
1MΩ?
Light up in red
no Convalue will
be repeated on the LCD.
Light up in red
no LnEvalue will
be repeated on the LCD.
Value (1). and (3). will be
alternately displayed on LCD
When the resistance between
LN is about 100kΩ or more
no OFF is displayed
alternately on the LCD.
No
No
No
Yes
Yes
Yes
PASS
Light up in green
Light up in green
Is the Switch
turning on?