Page view 14
13
   Press Null|250V/500V switch with contacting the Test Lead and the
metal parts, the resistance of Test Lead will be displayed on the LCD as
shown above fig.10 for 2sec.
   Then, the instrument cancels the resistance value of Test Lead and
adjust the displayed value to 0.00 as shown below fig.10.
   At this bout,
NULL mark is displayed in the LCD.
   Null setting cannot be done when a resistance is 3Ω or more.
   A message no appears to indicate a resistance is exceeding the Null
setting range.
Display at Null setting
 ⑷ Null setting can be released by pressing Null|250V/500V
switch for
2sec. The
NULL mark on the LCD will disappear when Null setting is
released.
   Null setting and release can be done on Class I Test function only.
5.2.4 Voltage setting for insulation resistance measurement
   (How to change 250V and 500V)
 ⑴ Set the function switch to Class II Test function, and press the
Null|250V/500V
switch. Then the mark to indicate the selected
voltage is shown on the LCD. By pressing Null|250V/500V
switch,
250V and 500V can be changed over.
Fig. 10
Fig. 11