16
7When testing is complete release the test button BEFORE disconnecting
the test leads from the circuit or from the appliance. This will ensure that
the charge built up by the circuit or the appliance during insulation test is
dissipated in the discharge circuit. In the discharging process, an LED
illuminates and the live circuit warning buzzer will sound.
CAUTION
NEVER TURN THE FUNCTION DIAL WHILST THE TEST BUTTON
IS DEPRESSED AS THIS MAY DAMAGE THE INSTRUMENT.
N EVER TOUCH THE CIRCUIT, TEST LEAD TIPS OR THE
APPLIANCE UNDER TEST DURING INSULATION TESTING.
NoteIf the reading measured greater than 200MΩ the over range reading
OLwill be displayed.