5. Specification
5.1 Measurement Specification
LoopImpedance
Function
(Operating Voltage)
Rated Voltage
Range
(Auto-Ranging)
NominalTestCurrent
at0ΩExternalLoop:
Magnitude/Duration
(
1)
Accuracy
Guaranteed Voltage
Range
L-PE
ATT OFF
(100
280
)
(45
65 Hz)
230V(50
60Hz)
L-PELOOP:
20Ω:0.00-19.99Ω
200Ω:20.0-199.9Ω
2000Ω:200-1999Ω
PFC/PSC:
2000A
0-1999A
20kA
2.00-19.99kA
L-PE:
20Ω
:
6A/20ms
200Ω
:
2.3A/20ms
2000Ω
:
15mA/250ms
L-N:
6A/20ms
±(3%rdg+4dgt)
(
2)
230 V (+10%/-15%)
(50/60 Hz)±1%
ATT ON
(100
280
)
(45
65 Hz)
230V(50
60Hz)
L-PELOOP:
20Ω:0.00-19.99Ω
200Ω:20.0-199.9Ω
2000Ω:200-1999Ω
PFC/PSC:
2000A
0-1999 A
20kA
2.00-19.99 kA
(L-N<20Ω)
L-N:6A/60ms
N-PE:10mA
/approx. 5s
±(3%rdg+6dgt)
(
2)
230 V (+10%/-15%)
(50/60 Hz)±1%
L-N/L-L
(100
500
)
(45
65 Hz)
L-N
230V(50/60Hz)
L-L
400V(50/60Hz)
L-N/L-LLOOP:
20Ω:0.00-19.99Ω
PSC:
2000A
0-1999 A
20kA
2.00-19.99 kA
20Ω:6A/20ms
L-N
±(3%rdg+4dgt)
L-L
±(3%rdg+8dgt)
(
3)
L-N
230 V (+10%
-15%)
L-L
400V(+10%
-15%)
(50
60 Hz)±1%
1: at 230V
2:Accuracy of L-N LOOP displayed on the Sub Display is
synchronizedwiththeoneatL-N/L-Lfunction.
  
PSC/PFCAccuracyisderivedfrommeasuredloopimpedance
specification and measured voltage specification.
3:PSCAccuracy is derived from measured loop impedance
specification and measured voltage specification.